Instructions for use of YXQ-LS-50A pressure steam sterilizer

This sterilizer is a device that uses saturated steam to sterilize medical devices, dressings, liquid medicine, glassware, etc. It has the advantages of excellent effect, easy operation, safety and reliability, and is suitable for medical treatment, printing and dyeing, scientific research, food health station and other units, and is an ideal sterilization equipment.

2. Structural characteristics

★ The shape is square vertical structure. The shell, sterilizer barrel and sterilization net basket are made of stainless steel.

★ Equipped with safety valve, when the pressure in the barrel reaches 0.25Mpa, the safety valve can automatically jump up, release the excessive pressure, and the performance is reliable.

★ Equipped with dual scale pressure gauge, with pressure and temperature reading indication, clear observation.

★ The container has an automatic protection function. When the container is dry and dry, it will automatically cut off the heater circuit and stop the sterilizer. It effectively protects the use of the equipment. At the same time, the buzzer sounds and the instrument displays Err3 to remind the user to replenish the water in time. .

★ New intelligent digital display regulator with single-chip technology for temperature and time. The user can set and adjust the temperature and time by operating the touch keys on the control panel according to different requirements to achieve the purpose of sterilization.

3. Technical parameters

a. Sterilizer volume: 50L; b. Temperature control range: 109 ℃ ~ 135 ℃;

c. Power supply voltage: AC 220V ± 10%, 50HZ ± 1%;

d. Heater power: 2.8KW;

e. Maximum working pressure: 0.212MPa; temperature: 135 ℃;

f. Ambient temperature: 5 ℃ -40 ℃; relative humidity≤93% RH.

4. Installation requirements

★ Please refer to this manual before use.

★ This equipment is classified as Class Ⅰ according to electric shock protection, and cannot be used in flammable and explosive anesthetic gas environment.

★ Warning "Steam is not exhausted, do not open": After the sterilization of the sterilizer is completed, the container cover must not be opened before the pressure gauge has returned to zero, otherwise it may cause high temperature steam burns.

★ This equipment is a fixed installation equipment. In order to ensure safe use, the user must install a fuse or overcurrent protector in the power circuit. The capacity of the fuse is 25 amps.

★ The user connects the equipment to the specified power supply according to the circuit wiring diagram according to the parameters specified in the product nameplate and instruction manual, and the equipment should be reliably grounded.

5. How to use

A stacked:

1. Rotate the handwheel (1) counterclockwise, pull the bucket cover (4) to the right, remove the sterilization basket (6), and remove the water baffle (7).

2. Open the small door (15), close the water release valve, add fresh water into the sterilization barrel (9), and the water level can reach the foot of the sterilization barrel (8) (under the water barrier). For continuous use, the amount of water must be made up after each sterilization.

3. Put back the water baffle and put the sterilization net basket into the sterilization barrel. Sterilized items are properly wrapped and placed in a sterilized mesh basket in order. There is a gap between each other, which is conducive to the penetration of steam and improves the sterilization effect.

B seal

1. Push the lid to the left and align it with the mouth of the bucket.

2. Check carefully and observe that the lid is aligned with the mouth of the sterilization drum, and then tighten the handwheel clockwise until it cannot be turned, so that the lid seal ring is tightly closed with the mouth of the drum, and the self-locking device (3 ) Coincides with the gear recess.

3. Connect the rubber tube to the steam vent connection on the upper part of the water storage barrel, and then insert it into a water storage barrel (14) containing about 100ml of cold water, and close the manual steam release valve (2). (Close clockwise and open counterclockwise) During the heating process, when the temperature controller displays a temperature of less than 102.0 ° C, the solenoid valve controlled by the temperature controller will automatically release steam to exclude the cold air in the sterilization barrel. When the display temperature is greater than 102.0 ° C, the automatic steam release stops. At this time, if a large amount of steam is still released, the manual steam release valve is not closed tightly, and it should be closed in time.

c heating:

1. Turn on the power switch to turn on the power, the digital display controller displays. At this point, you can start to set the temperature and sterilization time.

2. Setting method:

① Press the “Set” button, and the temperature setting value (℃) appears.

â‘¡ Press the "Set" button again, it appears: Set the timer time (minutes)

③ Press the “Set” button again, it appears, this state is related to temperature correction, enter the password, press the Set button to enter the correction temperature value: display, at this time, press ▲ ▼ to modify the temperature correction value. If the password is not entered or the password is incorrect, go to ④.

④ Press the “Set” button again, it appears, and you can set the steam release mode at this time. There are three options for the steam release mode. Mode 1: The steam release solenoid valve does not operate after sterilization without exhausting steam. Mode 2: Pulse steam exhaust means that the steam release solenoid valve is opened and closed when the sterilization is completed, and the steam is exhausted in the form of pulse. Mode 3: Normal exhaust steam means that the steam release solenoid valve keeps opening the exhaust steam after sterilization. The above three return to the temperature display to complete the setting.

⑤ Press the "work" button, the "work" indicator lights, the system works normally, and enter the automatic control sterilization process.

â‘¥If it is burnt dry, the switch is automatically disconnected, and there is an alarm sound.

⑦ After the work is completed, you must wait for the sterilizer pressure to drop below 107 ℃ before allowing the sterilizer door to open.

⑧Other key functions:

"" Key: In the state of setting parameters, each time it is pressed, the set parameter moves to the left. In the setting state, press "â–²" "â–¼" key: increase / decrease the value of parameter setting bit.

"Switch" key "During the measurement process, switch to display the current temperature value and the remaining time value.

Steam release button: Each press of the steam release button opens / closes the steam release solenoid valve.

"Time" light

When the temperature of the inner cavity reaches the set temperature, the timer light will turn on and the sterilization will start to time.

When the temperature of the inner cavity is lower than the set temperature by 0.5 ℃ or less, the timer light will be off and no time will be counted.

When the sterilization time reaches the set time, the sterilization is completed. The timer light is off, the work light is off, and ten beeps sound, and the panel displays "End" to end the work.

"Alarm" light

When the temperature of the inner cavity is higher than the set temperature by more than 2 ℃, the alarm light is on, the panel displays “Er2”, and the alarm sounds at the same time.

When the temperature of the inner cavity drops back to less than the set temperature +1.5 ℃, the alarm light is off and the alarm sound is turned off.

Note:-* The parameters in this step and the following steps are related to the working performance of the device, users should not change or set it easily. If the user really needs to adjust the parameters, he can follow the steps below.

â‘  When the parameter appears, if the parameter is set to

Press the "Settings" key to appear and the zero correction can be modified at this time.

a. If the actual temperature is 110.0 ° C and the display temperature is 109.0 ° C, the correction value = 110.0-109.0 = 1 ° C.

b. If the actual temperature is 113.2 ° C and the display temperature is 115.0 ° C, the correction value = 113.2-115 = -1.8 ° C. (Ie: correction value = actual temperature-display temperature)

D sterilization

1. The setting pressure of the safety valve of this instrument is 0.25MPa.

2. When the set temperature and sterilization time are completed, the electronic control device will automatically turn off the heating power, the "work" indicator "time" indicator is off, and accompanied by a beep to remind, the panel displays "End", at this time End of sterilization.

4 After the sterilization is completed, the steam exhaust solenoid valve will work in three modes.

Modes: no exhaust, pulse exhaust and normal exhaust, the three working modes can be manually intervened (press the steam release button). And at any time when the autoclave is working, you can press the steam release button to interfere with the steam release / no steam release of the sterilizer.

Sixth, attention and maintenance

★ In the use of the equipment, the safety valve should be maintained and inspected. When the equipment is idle for a long time and reused, the small wrench on the safety valve should be pulled to check whether the valve core is flexible to prevent the safety valve from jumping due to the corrosion of the spring.

At the same time when the equipment is working, when the pressure gauge indicates more than 0.25Mpa, the safety valve does not open, you should immediately turn off the power, open the vent valve knob, when the pressure gauge pointer returns to zero, wait for 1-2 minutes, then open the container Replace the safety valve in time.

★ The pressure gauge should be verified within the prescribed time limit to ensure safe use. In daily use, if the pressure gauge indicates unstable or cannot return to zero, it should be repaired or replaced with a new gauge.

★ The rubber sealing ring on the container cover of this equipment will age and deform over time. In addition to proper use in daily use, if the sealing ring is deformed or broken, it should be replaced in time to ensure safe use.

★ If loose screws or nuts are found during daily use of this equipment, it should be fastened in time to ensure normal use.

★ When stacking sterilized items, it is strictly forbidden to block the air outlet of the safety valve (12), and a space must be left to ensure its smooth deflation.

★ Before each use, you must check whether the amount of water in the barrel is kept at the foot of the sterilization barrel.

★ When the sterilizer continues to work, when new sterilization is carried out, it should be left for 5 minutes, and the upper cover is opened to allow the equipment to cool down.

★ When sterilizing the liquid, the liquid canister should be packed in a hard heat-resistant glass bottle, preferably no more than 3/4 volume. The bottle mouth is made of cotton gauze, and do not use unopened rubber or cork. Special attention: the steam is not allowed to be released immediately at the end of the sterilization liquid. The residual steam must be discharged after the pressure gauge pointer returns to zero.

★ For items of different types and different sterilization requirements, such as dressings and liquids, do not put them together for sterilization, so as to avoid loss and loss.

★ Normally keep the equipment clean and dry before extending the service life.

★ This equipment is a special equipment for pressure steam sterilization, which should not be used for other purposes.

Automatic vertical pressure steam sterilizer